N2 - A two-phase argon detector is being developed in our laboratory for dark matter search and low-energy neutrino experiments. A 109Cd γ-ray source was used to calibrate the energy scale of the detector. In this paper, a detailed emission spectrum of the 109Cd source was measured using a YAP:Ce based scintillation detector and high-purity germanium detector. It was shown that the 109Cd source equipped with a tungsten substrate and a copper filter can provide a complete set of γ-ray lines in the range from 8 to 90 keV for energy calibration of two-phase detectors. These measurements allowed us to successfully quantify the shape of the amplitude spectrum observed when the two-phase detector was irradiated by the 109Cd source.
AB - A two-phase argon detector is being developed in our laboratory for dark matter search and low-energy neutrino experiments. A 109Cd γ-ray source was used to calibrate the energy scale of the detector. In this paper, a detailed emission spectrum of the 109Cd source was measured using a YAP:Ce based scintillation detector and high-purity germanium detector. It was shown that the 109Cd source equipped with a tungsten substrate and a copper filter can provide a complete set of γ-ray lines in the range from 8 to 90 keV for energy calibration of two-phase detectors. These measurements allowed us to successfully quantify the shape of the amplitude spectrum observed when the two-phase detector was irradiated by the 109Cd source.
UR - http://www.scopus.com/inward/record.url?scp=85077062093&partnerID=8YFLogxK
U2 - 10.1134/S0020441219050142
DO - 10.1134/S0020441219050142
M3 - Article
AN - SCOPUS:85077062093
VL - 62
SP - 746
EP - 749
JO - Instruments and Experimental Techniques
JF - Instruments and Experimental Techniques
SN - 0020-4412
IS - 6
ER -
ID: 22981357