Результаты исследований: Научные публикации в периодических изданиях › статья › Рецензирование
A multichannel neutron flux monitoring system for a boron neutron capture therapy facility. / Bykov, T. A.; Kasatov, D. A.; Koshkarev, A. M. и др.
в: Journal of Instrumentation, Том 14, № 12, P12002, 05.12.2019.Результаты исследований: Научные публикации в периодических изданиях › статья › Рецензирование
}
TY - JOUR
T1 - A multichannel neutron flux monitoring system for a boron neutron capture therapy facility
AU - Bykov, T. A.
AU - Kasatov, D. A.
AU - Koshkarev, A. M.
AU - Makarov, A. N.
AU - Porosev, V. V.
AU - Savinov, G. A.
AU - Shchudlo, I. M.
AU - Taskaev, S. Y.
PY - 2019/12/5
Y1 - 2019/12/5
N2 - The progress in the design of new accelerator-based neutron sources for boron neutron capture therapy (BNCT) generated new requirements to the detector systems to monitor the particle flux during the irradiation of patients. In the current research, we evaluated basic parameters of a twin scintillator-over-fiber system with a silicon photo-multiplier readout based on boron-enriched and boron-free plastic scintillators. The presented results showed that the proposed system could be used for monitoring of neutron flux, and the designed detector system demonstrated good linearity, up to ∼ 2 mA, limited by accelerator specifications at that time.
AB - The progress in the design of new accelerator-based neutron sources for boron neutron capture therapy (BNCT) generated new requirements to the detector systems to monitor the particle flux during the irradiation of patients. In the current research, we evaluated basic parameters of a twin scintillator-over-fiber system with a silicon photo-multiplier readout based on boron-enriched and boron-free plastic scintillators. The presented results showed that the proposed system could be used for monitoring of neutron flux, and the designed detector system demonstrated good linearity, up to ∼ 2 mA, limited by accelerator specifications at that time.
KW - Detector modelling and simulations I (interaction of radiation with matter, interaction of photons with matter, interaction of hadrons with matter, etc)
KW - Instrumentation for hadron therapy
KW - Neutron detectors (cold, thermal, fast neutrons)
KW - Scintillators, scintillation and light emission processes (solid, gas and liquid scintillators)
KW - SCINTILLATOR
KW - DOSIMETRY
UR - http://www.scopus.com/inward/record.url?scp=85080102046&partnerID=8YFLogxK
U2 - 10.1088/1748-0221/14/12/P12002
DO - 10.1088/1748-0221/14/12/P12002
M3 - Article
VL - 14
JO - Journal of Instrumentation
JF - Journal of Instrumentation
SN - 1748-0221
IS - 12
M1 - P12002
ER -