Research output: Chapter in Book/Report/Conference proceeding › Conference contribution › Research › peer-review
Development, calibration and application of new-generation dissector with picosecond temporal resolution. / Meshkov, O. I.; Anchugov, O. V.; Dorokhov, V. L. et al.
Proceedings of the 5th International Beam Instrumentation Conference, IBIC 2016. Joint Accelerator Conferences Website (JACoW), 2016. p. 205-208 (Proceedings of the 5th International Beam Instrumentation Conference, IBIC 2016).Research output: Chapter in Book/Report/Conference proceeding › Conference contribution › Research › peer-review
}
TY - GEN
T1 - Development, calibration and application of new-generation dissector with picosecond temporal resolution
AU - Meshkov, O. I.
AU - Anchugov, O. V.
AU - Dorokhov, V. L.
AU - Kurkin, G. Ya
AU - Petrozhitsky, A. N.
AU - Shvedov, D. V.
AU - Zinin, E. I.
AU - Gornostaev, P. B.
AU - Schelev, M. Ya
AU - Shashkov, E. V.
AU - Smirnov, A. V.
AU - Zarovskii, A. I.
PY - 2016/1/1
Y1 - 2016/1/1
N2 - A dissector is an electron-optical device designed for measurement of periodic light pulses of subnanosecond and picosecond duration. LI-602 dissector developed at BINP SB RAS is widely used for routine measurements of a longitudinal profile of electron and positron beams at BINP electron-positron colliders and other similar installations [1-3]. LI-602 dissector is a part of many optical diagnostic systems and provides temporal resolution of about 20 ps. Recently a new generation of picosecond dissectors were created on the basis of the PIF-01/S1 picosecond streak-image tube designed and manufactured at the GPI Photoelectronics Department [4, 8-10]. The results of the measurements of instrument function of the new dissector based on PIF-01/S1, which were carried out in the static mode [5], showed that temporal resolution of the dissector can be better than 3-4 ps (FWHM). The results of temporal resolution calibration of the new-generation picosecond dissector carried out at the specialized set-up based on a femtosecond Ti:sapphire laser and recent results of longitudinal beam profile measurements at BINP damping ring are given in this work.
AB - A dissector is an electron-optical device designed for measurement of periodic light pulses of subnanosecond and picosecond duration. LI-602 dissector developed at BINP SB RAS is widely used for routine measurements of a longitudinal profile of electron and positron beams at BINP electron-positron colliders and other similar installations [1-3]. LI-602 dissector is a part of many optical diagnostic systems and provides temporal resolution of about 20 ps. Recently a new generation of picosecond dissectors were created on the basis of the PIF-01/S1 picosecond streak-image tube designed and manufactured at the GPI Photoelectronics Department [4, 8-10]. The results of the measurements of instrument function of the new dissector based on PIF-01/S1, which were carried out in the static mode [5], showed that temporal resolution of the dissector can be better than 3-4 ps (FWHM). The results of temporal resolution calibration of the new-generation picosecond dissector carried out at the specialized set-up based on a femtosecond Ti:sapphire laser and recent results of longitudinal beam profile measurements at BINP damping ring are given in this work.
UR - http://www.scopus.com/inward/record.url?scp=85071956147&partnerID=8YFLogxK
M3 - Conference contribution
T3 - Proceedings of the 5th International Beam Instrumentation Conference, IBIC 2016
SP - 205
EP - 208
BT - Proceedings of the 5th International Beam Instrumentation Conference, IBIC 2016
PB - Joint Accelerator Conferences Website (JACoW)
T2 - 5th International Beam Instrumentation Conference, IBIC 2016
Y2 - 11 September 2016 through 15 September 2016
ER -
ID: 21488985